Setting up the SmartGuard feature


Device: MiniMed™ 780G insulin pump

The SmartGuard™ feature requires a 48-hour warm-up period before activation. This warm-up period begins at midnight after the pump starts delivering insulin and it does not require sensor use. During the warm-up period, the pump collects and processes data for use by the SmartGuard™ feature.

Once turned on, SmartGuard™ technology begins automatically when all conditions are met and a blood glucose (BG) is entered.


To turn the SmartGuard feature on or off:

  1. From the Home screen, press Select.

  2. Select SmartGuard.

    Select SmartGuard screen

  3. Select SmartGuard to turn the feature on or off.

    Select SmartGuard on or off screen

Note: Certain additional requirements must be met before the SmartGuard feature activates. See the SmartGuard Checklist for more information.


To confirm or change the SmartGuard settings:

  1. From the Home screen, press Select.

  2. Select SmartGuard.

    Select SmartGuard screen

  3. Select SmartGuard Settings.

    Select SmartGuard Settings screen

  4. Select Target to adjust the SmartGuard target: 100 mg/dL (default), 110 mg/dL, or 120 mg/dL.

    Select Target screen

    Note: The Auto correction feature is on by default.

  5. Select Save after making any changes..