SmartGuard Checklist


Device: MiniMed 780G insulin pump

The SmartGuard Checklist screen indicates what’s needed to start or continue using the SmartGuard feature.

To use the SmartGuard Checklist:

  1. From the Home screen, press Menu select, and then select Menu shield.
  2. Select SmartGuard Checklist.

To check the SmartGuard Checklist


SmartGuard Checklist table


Line If this appears Do this
1 To check the SmartGuard Checklist Enter a new blood glucose (BG) meter reading.
To check the SmartGuard Checklist
To check the SmartGuard Checklist The system requires a BG reading and will ask when it is ready.
2 Turn on SmartGuard in the SmartGuard menu Turn on SmartGuard in the SmartGuard menu.
3 To check the SmartGuard Checklist

a) Confirm the pump shows a sensor serial number on the Paired Devices screen. Make sure the pump is paired with a sensor.

b) Check your Home screen. If you see the connection icon with a red cross move your pump and transmitter closer together. The pump will try to find the transmitter signal which may take 15 minutes. If after 30 minutes the pump and transmitter are still not communicating, you will receive a Lost sensor signal alert. Check that the sensor is still inserted in the skin and the transmitter and sensor are still connected. Move your pump closer to your transmitter.

c) If your sensor glucose is outside of the 50 to 400 mg/dL range, your pump will not enter the SmartGuard feature.

To check the SmartGuard Checklist Turn on the sensor in Settings > Device Settings.
To check the SmartGuard Checklist Pair the pump and sensor.
4 Bolus in progress menu Wait until the bolus is complete or stop the bolus yourself before the SmartGuard feature can be used.
5 Delivery suspended menu If insulin delivery is suspended, the SmartGuard feature cannot be used. Treat low BG if necessary, as instructed by your healthcare professional.
6 Carb ratio not set menu When you turn on the Bolus Wizard feature for the first time, enter your Carb Ratio in the Edit Carb Ratio screen. You can also enter your Carb Ratio in the Bolus Wizard Setup screen, even if the Bolus Wizard feature is not turned on.
7 Temp Basal rate menu If a temp basal is currently active, you must wait until it has completed or cancel the temp basal yourself before the SmartGuard feature can be used.
8 SmartGuard updating menu If SmartGuard active insulin is currently updating, it will take up to 5 hours to complete. You must wait until this amount is updated before the SmartGuard feature can be activated.
9 SmartGuard warming up menu Wait for the SmartGuard feature to gather insulin delivery history and determine the basal rate.