Change a Basal Rate
Device: MiniMed™ 670G insulin pump (MMT-1780K)
Please note that the basal rates used below are for illustration purposes only. Your basal settings will be different. Your healthcare provider will give you the appropriate basal rate.
To change a basal rate:
1. From the Home screen, press Select.
2. Select Basal.
3. Select Delivery Settings.
4. Select Basal Pattern Setup.
5. Select the Basal Pattern you wish to edit.
6. Select Options.
7. Select Edit.
8. Press select on the time segment.
9. Press select on End time.
10. Press up or down to change U/hr value and press Select.
11. Select Done.
12. Review rates and select Save.