Programming a Preset Temp Basal rate


Device: MiniMed 780G insulin pump

The Preset Temp feature lets you setup basal rates for recurring short-term situations where you need to temporarily change your basal rate. There are four names you can use to match your preset temp basal rate to a situation: High Activity, Moderate Activity, Low Activity, and Sick. There are also 4 additional preset temp rates available to use for other circumstances (Temp1 through Temp 4). A preset temp basal can be a good option for postpartum women, patients who are on steroids, or patients who recently had bariatric surgery.

  1. From the Home screen, press Select.

  2. Select Insulin.

    Select Insulin screen

  3. Select Basal.

    Select Programming a Preset Temp Basal rate screen

  4. Select Delivery Settings.

    Select Programming a Preset Temp Basal rate screen

  5. Select Preset Temp Setup.

    Select Programming a Preset Temp Basal rate screen

  6. Select Add New.

  7. Select a name for the Preset Temp Basal rate.

    Select Programming a Preset Temp Basal rate screen

  8. Select Type to choose between Percent and Rate.

    Select Programming a Preset Temp Basal rate screen

  9. Enter in a percentage, if you choose Percent or a rate if you choose Rate. Press Select.

    Select Programming a Preset Temp Basal rate screen

  10. Set the Duration using the Up and Down buttons. Press Select.

    Select Programming a Preset Temp Basal rate screen

  11. Select Save.

Learn how to start a Preset Temp Basal rate on your pump when you need to use it.