Alert Silence
Device: MiniMed™ 770G insulin pump
The Alert Silence option lets you temporarily silence sensor glucose (SG) alerts. This is useful in situations where you do not want to disturb others, such as when you are in a business meeting or in a movie theater. When using this option, your system still records the time and glucose value for any alerts that occur. You can view this information in the Alarm History screen.
If a glucose alert occurs when you are using the Alert Silence option, the notification light begins to flash and the Sensor alert occurred alert appears to let you know an alert was silenced, but there is no vibration or sound. If you have not cleared the alert by the end of the preset alert silence duration, your pump begins to beep and/or vibrate periodically until the alert is cleared.
To set Alert Silence in Auto Mode:
Press select, select Audio Options and select Alert Silence Options.
- Select High Alerts Only, High and Low Alerts, or All Sensor Alerts to set the alerts you want silenced. Refer to the table below for details about the alerts silenced with each selection.
- Set the Duration. The duration can be set in 30-minute increments from 30 minutes to 24 hours.
- Select Begin. The Alert Silence settings immediately take effect and you are returned to the Sensor Settings screen.
The following table describes the glucose alerts that are silenced with each option.
Option | Silences these alerts |
---|---|
High Alerts Only | Alert on high, Alert before high, and Rise Alert |
High & Low Alerts | Alert on high, Alert before high, Rise Alert, Alert on low, and Alert before low |
All Sensor Alerts | All of the alerts listed previously for High and Low Alerts, plus the following:
|
To cancel Alert Silence:
- Press select, select Audio Options and select Alert Silence Options.
Select Cancel Alert Silence.